NANOLITER 2020/200704
World Precision Instruments
33
APPENDIX B: MICRO2T COMPUTER CONTROL
Serial commands are used to control the
MICRO2T
via the serial port of a computer using
a free USB port.
Serial Commands
All commands are case sensitive. The settings for the serial port are 9600 baud rate, 8 data
bits, 1 start bit and 1 stop bit. Numbers and decimal points are indicated below by the “#”
symbol. Enter a carriage return (
Enter
key on the keyboard) after each command.
Command
Syntax
Notes
Set target volume
V######## ######## - Desired Volume Target in nL.
Value can be entered with or without decimal
point (Range: 0 to 999999.9)
Set delivery rate
R######## ######## - Desired Delivery Rate in nL/
sec or nL/min. Units are set with S and M
commands. Value can be entered with or
without decimal point (Range: 0 to 999999.9)
Set active pump to infuse direction
I
Set active pump to withdraw direction
W
Start delivery from stopped condition or
resume from paused
G
Halt volume delivery (STOP)
H
Cancels delivery in process
Pause delivery
U
Pauses delivery in process
Set delivery units to nL/sec
S
Set delivery units to nL/min
M
Set active pump
L#
# is between 1 and 2
Set mode to non-grouped
N
Set mode to grouped
P
Set mode to disabled
D
Select syringe
T#
# Is between 1 and 12. 1–9 for types 1–9,
10–12 for A, B and C.
Set motor drive to max load
BT
Set motor drive to smooth
BS
Set volume counter mode to Delivered
EN
Set volume counter mode to Remaining
EI
Display target volume
?V
Response: Target Volume = 10000.0nL
OK
Display volume counter
?C
Response: Volume Counter = 1000.0µL
OK
Display motor drive option
?B
Response: Smooth Drive (Max Load Drive)
OK
Display motor counter mode
?E
Response: Delivered Volume
(Remaining Volume)
OK
Display pump mode
?M
Response: Mode: Non-Grouped
(Mode: Grouped, Mode: Disabled)
OK