NANOLITER 2020/200704
World Precision Instruments
15
3.
Volume Counter
determines how the volume counter is affected when volume is
delivered. Tap the control to toggle between
Delivered
and
Remaining
.
•
When the
Volume Counter
slider is set to
Delivered
, the volume counter
increments when injecting and decrements when withdrawing. The volume count
shown is the amount delivered.
•
When
Run
is pressed the counter resets to zero and increments. When the
slider is set to
Remaining
, the volume counter decrements when injecting and
increments when withdrawing. When
Run
is pressed the counter is set to the
volume target and decrements. Negative values signify a withdrawal and positive
indicate an infusion.
4. Tap on a round button to select the desire units. Choose between nL/sec and nL/min
for the delivery rate.
5. Tap on a round button to select your mode. Mode buttons determine the pump
operating mode.
•
Disabled
–The selected pump is disabled and will not operate. When you return to
the Command screen, DISABLED appears over the selected channel.
•
Grouped
–You can start and stop all grouped pumps simultaneously. See
“Grouped Operations” on page 32. When
Grouped
is selected, both pumps (if
present) are set to grouped mode.
•
Non-Grouped
–Non-grouped pumps operate independently.
Calibrating the MICRO2T System Travel Limits
The
MICRO2T
monitors the location of the
NANOLITER2020
plunger. It is important to
define the limits of travel before you begin using the pump. This keeps the pump from
driving past the mechanical limits of the pump and avoids any jamming issues. When you
install a pump on a new channel, you must set the stop positions so the controller will
properly monitor the pump. Usually this only needs to be done once. But, if the pump is
moved to another channel or the pump, you will need to reset the end stops.
1.
End Stop
: The
End Stop
calibration procedure sets the limit for the maximum
distance that the pump’s plunger carrier can travel in the withdrawn direction. This
setting prevents the motor from forcing the plunger carrier to the mechanical limit
of the drive mechanism for an extended period of time, in the event the pump is left
running unattended. This prevents the possibility of binding the lead screw which can
cause the pump to lock up. The ideal location for the
End Stop
limit is just a fraction of
a millimeter before the maximum mechanical limit is reached.
The calibration procedure is simple and involves using the controller to manually
position the
NANOLITER2020
into the fully withdrawn position by pressing a button.
When the
END STOP
button is released, the controller automatically performs the
calibration. The calibration should be checked:
•
Every time a pump is newly connected to a
MICRO2T
controller
•
Any time a pump is switched to a new channel that was not previously calibrated.
•
Any time a pump is stalled due to excessive load or other accidental misuse.