WIKA operating instructions, model CPH6200
EN
11221780.02 04/2019 EN/DE
6. Commissioning, operation
6.4.7 Power-off delay (
P
.
oFF
)
If no button is pressed and no serial communication occurs during the power-off delay,
the instrument will automatically switch itself off. The power-off delay can be set between
1 and 120 min. If
“
P
.
oFF
=
oFF
”
then the power-off function is deactivated.
6.4.8 Instrument output (
Ovt
)
The output can be used as a USB or RS-232 interface or as an analogue output (0 ... 1 V).
6.4.9 Analogue output scaling with
dAC
.
0
and
dAC
.
1
(
dAC
.)
WARNING!
Damage to property through incorrect measuring instruments
By using incorrect measuring instruments, this damage could occur to the
hand-held.
▶
Only connect passive voltmeters to the analogue output.
Analogue output cannot be used during logger recordings.
With
dAC
.
0
and
dAC
.
1
the analogue output can be easily scaled.
▶
It must be ensured that the analogue output is not loaded too heavily, otherwise the
output value can be corrupted and the current supply of the instrument increases
correspondingly.
Loads up to approx. 10 kΩ are harmless.
⇒
If the display exceeds the value set with
dAC
.
1
, then 1 V will be output.
⇒
If the display drops below the value set with
dAC
.
0
, then 0 V will be output.
⇒
In the event of an error (
Err
.
1
,
Err
.
2
, ----, etc.) an analogue signal slightly over 1 V will
be output.
Jack connector wiring
WARNING!
Damage to property through incorrect jack connector
Using an incorrect jack connector or incorrect wiring can cause damage to
the hand-held.
▶
The 3rd connection must not be used.
▶
Only stereo jack connectors are permitted.
▶
Only use the original connection cable from WIKA.