WIKA operating instructions, model CPH6200
EN
11221780.02 04/2019 EN/DE
6. Commissioning, operation
6.3 Menu functions
Menu Para-
meter
Values
Meaning
[MENU]
▶
▲
or
▼
SE
t
ConF
Set Configuration: General settings
Unit
mbar, bar, ...
Display of unit
1) 2)
SL
oFF
/
on
Sea level: Switch sea-level correction on or off
1) 2)
ALti
-200 ... +9999
Altitude: Input of altitude above sea level [metre] (only if SL is
activated)
1) 2)
rAtE
Rate: Measuring rate (see chapter 6.4.3 “Measurement types (rAtE)”)
1)
SLo
Slow: Slow measurement (4 Hz filtered, low current
consumption)
1)
FASt
Fast: Fast measurement, filtered (> 1,000 Hz)
1)
P
.
dEt
Peak detection: Fast measuring rate, unfiltered (> 1,000 Hz)
1)
t
.
AU6
1 ... 120
Time in seconds, calculated via the averaging function
3)
oFF
Averaging deactivated
3)
P
.
oFF
1 ... 120
Auto power-off delay in minutes.
If no button is pressed and there is no data transfer via the
interface, the instrument will switch itself off after this interval.
oFF
Auto power-off function deactivated (continuous operation)
Ovt
oFF
No output function, lowest current consumption
SEr
Instrument output is serial interface
dAC
Instrument output is analogue output 0 ... 1 V
Adr
.
01, 11 ... 91
Communications address of the interface (only with
Out
=
SEr
)
dAC
.
CH1, CH2 or
CH DIF
Measuring input which should be used for the analogue output
(only with
Out
=
dAC
)
dAC
.
0
Sensor dep.,
e.g. -5.00 ...
+5.00 mbar
Zero point setting for
Out
=
dAC
: Input of the measured value
at which the analogue output should output 0 V (only for
Out
=
dAC
)
dAC
.
1
Sensor dep.,
e.g. -5.00 ...
+5.00 mbar
Scale setting for
Out
=
dAC
: Input of the measured value
at which the analogue output should output 1 V (only for
Out
=
dAC
)
1) If there is data in the logger memory, this menu item cannot be accessed. If these should be changed, the data must first be
deleted (see chapter 6.5 “Operation of the logger function”).
2) This menu can only be selected if an appropriate sensor is connected to connection 1. When using a second corresponding
sensor on connection 2 then the settings are adopted.
3) If the automatic recording is activated, this menu item cannot be accessed. If these should be changed, the cyclic data logger
must first be stopped (see chapter 6.5 “Operation of the logger function”).