WIKA operating instructions, model CPH6200
EN
11221780.02 04/2019 EN/DE
6. Commissioning, operation
Menu Para-
meter
Values
Meaning
[MENU]
▶
▲
or
▼
SEt
CAL
Set Calibration: Adjustment of sensors
3)
OFS
.
1
Sensor dep.,
e.g. -5.00 ...
+5.00 mbar
The zero point of sensor 1 will be displaced by this value
to compensate deviations of the probe or the measuring
instrument.
oFF
Zero point offset is deactivated (= 0.000)
SCL
.
1
e.g. -5.00 ...
+5.00 mbar
The measuring scale of sensor 1 will be changed by this factor
[%] to compensate deviations of the probe or the measuring
instrument.
oFF
Scale correction factor deactivated (= 0.000)
OFS
.
2
Sensor dep.,
e.g. -5.00 ...
+5.00 mbar
The zero point of sensor 2 will be displaced by this value to
compensate for deviations of the probe or the measuring
instrument.
oFF
Zero point offset is deactivated (= 0.000)
SCL
.
2
e.g. -5.00 ...
+5.00 mbar
The measuring scale of sensor 2 will be changed by this factor
[%] to compensate deviations of the probe or the measuring
instrument.
oFF
Scale correction factor deactivated (= 0.000)
SEt
AL
.
Set Alarm: Alarm function configuration
4)
AL
.
[1,2,DIF]
on
Alarm sensor is on; is indicated acoustically
no
.
So
Alarm sensor is on; is indicated acoustically
oFF
No alarm function
AL
.
Lo
[1,2,DIF]
Min-Range ...
... AL.Hi
Min. alarm limit (not when
AL.oFF
, sensor min. is the lower
display range limit of the connected sensor)
AL
.
Hi
[1,2,DIF]
AL.Lo ...
... Max-Range
Max. alarm limit (not when
AL
.
oFF
, sensor max. is the upper
display range limit of the connected sensor)
SEt
Lo66
Set Logger: Logger function configuration
1)
Func
CYCL
Cyclic: Logger function “cyclic logger”
Stor
Store: Logger function “individual value logger”
oFF
No logger function
CYCL
1
...
3600
Cycle time for cyclic logger [seconds]
Lo
.
Po
on
/
oFF
Low-power logger with very low current supply (only with
cyclic logger and slow measurement)
SEt
CLOC
Set Clock: Setting of real-time clock
CLOC
HH:MM
Setting the time hours:minutes
dAtE
TT.MM
Setting the date day.month
YEAr
YYYY
Setting the year
1) If there is data in the logger memory, this menu item cannot be accessed. If these should be changed, the data must first be
deleted (see chapter 6.5 “Operation of the logger function”).
2) This menu can only be selected if an appropriate sensor is connected to connection 1. When using a second corresponding
sensor on connection 2 then the settings are adopted.
3) If the automatic recording is activated, this menu item cannot be accessed. If these should be changed, the cyclic data logger
must first be stopped (see chapter 6.5 “Operation of the logger function”).
4) If an alarm function limit is crossed (over or under), this is signalled by a “hooting” and a beeping.