68
Functional verification
Welch Allyn Propaq LT Vital Signs Monitor
System
1 Indicators, start-up
Red, yellow, and green LEDs
Low, medium, and loud tones
5 Voltage, power supply
Within range
7 Pixels
Visual
10 Buttons
Visual
13 Charger state
8
ECG
5 Waveform
Present
Heart rate
80 bpm ± 3 bpm
6-8 Single lead fail
(repeat for each lead)
Lead Fail message for each
lead
12 Pacer indicator
Present
Respiration
5 Waveform (Lead 1)
Present
Heart rate
20 bpm ± 2 bpm
8 Waveform (Lead 2)
Present
Heart rate
20 bpm ± 2 bpm
SpO
2
- monitor only, Nellcor
4 Pulse rate @ 60 bpm
60 bpm ± 3 bpm
Saturation @ 75%
75% ± 3%
6 Pulse rate @ 200 bpm
200 bpm ± 3 bpm
8 Saturation @ 90%
90% ± 3%
SpO
2
- monitor only, Masimo
2 Pulse rate
61 bpm ± 3 bpm
Saturation
81% ± 2%
SpO
2
- monitor and sensor
Pulse rate @ 60 bpm
60 bpm ± 4 bpm
Saturation @ 94%
94% ± 4%
(continued below)
Test
Specification
Result
Pass NA
Fail
Содержание Propaq LT 802LT0N
Страница 1: ...Propaq LT Vital Signs Monitor Service Manual Software version 1 5X or later ...
Страница 4: ...iv Contents Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 8: ...4 Safety summary Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 28: ...24 Troubleshooting and repair Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 74: ...70 Functional verification Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 78: ...74 35 ml test volume Welch Allyn Propaq LT Vital Signs Monitor ...