Service Manual
Disassembly and reassembly
35
c.
Remove the rubber gaskets from the display module.
d.
Set aside the display module, protecting the display surface from damage.
4.
Remove the speaker.
REASSEMBLY:
Before removing the speaker, note the alignment of the speaker cable terminals for
rerouting the cable.
a.
Pry the speaker from the adhesive gasket securing it to the chassis.
b.
Remove any remaining gasket pieces from the front panel and discard them.
REASSEMBLY:
1. Install a new gasket.
2. Align the new speaker for proper cable routing.
3. Seat the speaker on the gasket.
Display surface
Speaker cable terminals
Speaker cable connector