Service Manual
Functional verification
57
10. Verify that the charge current is between 800 mA and 1100 mA and that the charger
state is 2.
11. Verify that the battery icon on the monitor is green with a lightning bolt.
12. Press
to start the NIBP pump.
13. Verify after several seconds of pumping that the charge current is less than 220 mA.
14. Verify after several seconds of pumping that V BAT is between 8.14 V and 8.55 V.
15. Press
to stop the NIBP pump.
16. Leave the monitor in the cradle until the status indicator on the cradle goes off.
17. Verify that the battery icon on the monitor display contains no lightning bolt.
18. Navigate to
Setup > Service > Battery
, press
, and verify that the charger state is
6
.
19. Remove the monitor from the cradle and let the monitor continue to run for a minute.
20. Install the monitor in the cradle and verify that the status indicator on the cradle is
briefly green and then goes off.
Note
Pump startup may be delayed until power up NIBP testing is completed.
Exit
NIBP
NIBP
ID: SET_CAOJ1T9L
ID: SET_CAOJ1T9L
3:00:06P
3:00:06P
Adult
Adult
Rm
Rm
Service
Service
ECG
ECG
SpO2
SpO2
USB
USB
Radio
Radio Software
Software
Display
Display
Battery
Battery
Voltages
Voltages
Buttons
Buttons
Battery
Battery
Show Info
Show Info
Exit
Exit
NIBP
NIBP
ID: SET_CAOJ1T9L
ID: SET_CAOJ1T9L
3:00:06P
3:00:06P
Adult
Adult
Rm
Rm
Service
Service
ECG
ECG
SpO2
SpO2
USB
USB
Radio
Radio Software
Software
Display
Display
Battery
Battery
Voltages
Voltages
Buttons
Buttons
Battery
Battery
Show Info
Show Info
Battery Information
Battery Information
SN
SN
V BAT
V BAT
7.83
7.83
044801
044801
Charge current
Charge current
Battery state
Battery state
Charger state
Charger state
3
0.84
0.84
2
Содержание Propaq LT 802LT0N
Страница 1: ...Propaq LT Vital Signs Monitor Service Manual Software version 1 5X or later ...
Страница 4: ...iv Contents Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 8: ...4 Safety summary Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 28: ...24 Troubleshooting and repair Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 74: ...70 Functional verification Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 78: ...74 35 ml test volume Welch Allyn Propaq LT Vital Signs Monitor ...