5
55
Functional verification
Functional verification overview
This section describes the procedure for a complete functional test to support
recommended preventive-maintenance schedules.
Equipment required
The following equipment is required for functional verification of a fully configured
monitor.
Commercially available general-purpose/medical test equipment
Item
Manufacturer part number/specification
Power supply
Variable, 0-28 VDC, 0.100A (minimum), with voltage and current
indicators (for 1mA current measurement)
Digital pressure meter
Netech Digimano 1000 or equivalent
AC withstand voltage (hi-pot) tester
Associated Research 3605 or equivalent
Hi-Pot Return Test Probe
Associated Research 38082 Return Retractor
Hi-Pot High Voltage Test Probe
Associated Research 38083 Safe-T Probe (High Voltage Test Gun)
ECG Simulator
Fluke 217A or Equivalent
SpO
2
simulator (to test the monitor and SpO
2
sensor) Fluke (Biotek) Index2 XL/XLFE or equivalent
Masimo SpO
2
Functional tester, Masimo SET
(to test the monitor only)
Masimo SET Tester 1795
Cable, Masimo SET, DB9-LNCS (4’, 10’, or 14’)
LNC-04-WA 4’ LNCS cable or
LNC-10-WA 10’ LNCS cable or
LNC-14-WA 14’ LNCS cable
Adapter cable, Masimo
1645, MAC-1 SpO2 adapter cable (LNOP-to-LNCS)
Nellcor SpO
2
Functional tester (to test the monitor only)
Nellcor SRC-MAX
Extension cable (required for SRC-MAX)
Nellcor DEC-8
Syringe, 60-ml, Slip-tip, Luer
BD (Becton, Dickinson) 309654 or equivalent
Содержание Propaq LT 802LT0N
Страница 1: ...Propaq LT Vital Signs Monitor Service Manual Software version 1 5X or later ...
Страница 4: ...iv Contents Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 8: ...4 Safety summary Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 28: ...24 Troubleshooting and repair Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 74: ...70 Functional verification Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 78: ...74 35 ml test volume Welch Allyn Propaq LT Vital Signs Monitor ...