28
Disassembly and reassembly
Welch Allyn Propaq LT Vital Signs Monitor
2.
Insert the flex cable into the connector.
3.
Slide the latching piece toward the connector body until it clicks into place.
Connector locations
Caution
Do not attempt to remove a flex cable until the ZIF latch has been
opened.
Caution
Never attempt to disconnect any cable by pulling on the cable wires.
Always disconnect cables by grasping and pulling only on the connector halves.
Connector
Connects with:
Connector
Connects with:
Main Board
Charging Cradle - Interconnect Board
J1
ECG (flex)
J1
Cradle power board (3-conductor)
J2
On/Off Switch
J2
Cradle power board (10-conductor)
J3
SpO
2
(flex)
JP1
External USB
J4
Display module
P1
Propaq LT Monitor
J5
BDM
Charging Cradle - Cradle Power Board
J6
PCMCIA (radio) card
J1
I/O panel (DC input)
J7
Button pad
J2
Cradle interconnect board (10-conductor)
J8
Battery pack
J4
Cradle interconnect board (3-conductor)
J9
Pump/bleed valve
Radio Card
J10
Dump valve
(flex)
Antenna
J11
Speaker
P2
Power/USB (flex)
Содержание Propaq LT 802LT0N
Страница 1: ...Propaq LT Vital Signs Monitor Service Manual Software version 1 5X or later ...
Страница 4: ...iv Contents Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 8: ...4 Safety summary Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 28: ...24 Troubleshooting and repair Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 74: ...70 Functional verification Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 78: ...74 35 ml test volume Welch Allyn Propaq LT Vital Signs Monitor ...