22
Troubleshooting and repair
Welch Allyn Propaq LT Vital Signs Monitor
3.
Simultaneously press
and
to access the NIBP characterization service screens. A
warning message appears: “NIBP characterization affects neonatal cuff detection.”
4.
Press
to continue. The first steps of the procedure appear:
5.
Follow the steps as presented on the screen.
6.
Press
to begin characterization.
When the characterization is complete, this screen appears:
7.
Verify the characterization. (
Note
The standard neonatal NIBP hose (008-0265-xx) can be effectively blocked with
firm thumb pressure over the fitting opening.
Note
In this procedure, ‘the select button’ refers to
.
WARNING
Always verify after characterization. Failure to verify a new
characterization could result in overinflation of a neonatal cuff.
without cuff to the Propaq LT.
without cuff to the Propaq LT.
2. Block the open end of the hose.
2. Block the open end of the hose.
3. With the hose completely blocked,
3. With the hose completely blocked,
Show Info
Show Info
NIBP:
NIBP:
Pump Test
Pump Test
HALL, ROBERT E.
HALL, ROBERT E.
3456187
3456187
12:41:32P
12:41:32P
Adult,
Adult,
Rm 239
Rm 239
Audible Alarms Suspended
Audible Alarms Suspended
Service:
Service: NIBP
NIBP
ECG
ECG SpO2
SpO2 USB
USB Radio
Radio Software
Software
Display
Display Battery
Battery Voltages
Voltages Buttons
Buttons
Characterize NIBP
Characterize NIBP
1. Connect a standard neonatal NIBP hose
1. Connect a standard neonatal NIBP hose
Exit
Exit
Select ( ) to initialize NIBP characterization.
Select ( ) to initialize NIBP characterization.
press the select button.
press the select button.
PROPER NEONATAL CUFF DETECTION.
PROPER NEONATAL CUFF DETECTION.
Failure to perform the verification can result in
Failure to perform the verification can result in
overinflation of neonatal cuffs.
overinflation of neonatal cuffs.
Show Info
Show Info
NIBP:
NIBP:
Pump Test
Pump Test
HALL, ROBERT E.
HALL, ROBERT E.
3456187
3456187
12:41:32P
12:41:32P
Adult,
Adult,
Rm 239
Rm 239
Audible Alarms Suspended
Audible Alarms Suspended
Service:
Service: NIBP
NIBP
ECG
ECG SpO2
SpO2 USB
USB Radio
Radio Software
Software
Display
Display Battery
Battery Voltages
Voltages Buttons
Buttons
Characterize NIBP
Characterize NIBP
THE VERIFICATION TEST ASSURES
THE VERIFICATION TEST ASSURES
Exit
Exit
NIBP Characterized.
NIBP Characterized.
Perform verification procedure now.
Perform verification procedure now.
Содержание Propaq LT 802LT0N
Страница 1: ...Propaq LT Vital Signs Monitor Service Manual Software version 1 5X or later ...
Страница 4: ...iv Contents Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 8: ...4 Safety summary Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 28: ...24 Troubleshooting and repair Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 74: ...70 Functional verification Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 78: ...74 35 ml test volume Welch Allyn Propaq LT Vital Signs Monitor ...