Service Manual
Functional verification
67
Checklist and test results report form
Use a copy of this 3-page form to track your progress through the validation tests.
Test
Specification
Result
Pass NA
Fail
Charging/LED operation
4 Input current, cradle
< 50 mA
Status, cradle powered,
monitor not inserted
Power indicator: ON
Status indicator: OFF
6 Status, power adapter
connected
Power indicator: ON
Status indicator: OFF
8 Status, monitor running with
discharged battery
Status indicator: GREEN
10 Charge current, battery fully
discharged
800 mA - 1100 mA
Charger state, battery fully
discharged
2
11 Status, battery charging
(green)
13 Charge current, pump running < 220mA
14 Battery voltage, pump running 8.14 V - 8.55 V
17 Status, battery fully charged
(green)
18 Charger state, battery fully
charged
6
20 Status, cradle transition
Status indicator:
green and then OFF
Battery charge time (optional)
2 Charge time
< 4 hours
(continued below)