June 28, 2016, 715003956 Rev. C
Page ix
Audience and purpose
This guide is for novice users, and assumes no knowledge of liquid chromatography or mass
spectrometry principles. It provides an overview of the instrument and explains how to install it,
prepare it for operation, and maintain it.
Intended use of the ACQUITY QDa Detector
Waters designed the ACQUITY QDa Detector for use as an ion confirmation and quantitation tool,
as part of an ACQUITY UPLC or UPC
2
system, Alliance, and LC and SFC prep systems. The
ACQUITY QDa Detector
is not intended for use in diagnostic applications.
Calibrating
To calibrate LC/MS systems, follow acceptable calibration methods using at least five standards to
generate a standard curve. The concentration range for standards should include the entire range
of QC samples, typical specimens, and atypical specimens.
Quality-control
Routinely run three QC samples that represent subnormal, normal, and above-normal levels of a
compound. If sample trays are the same or very similar, vary the location of the QC samples in the
trays. Ensure that QC sample results fall within an acceptable range, and evaluate precision from
day to day and run to run. Data collected when QC samples are out of range might not be valid. Do
not report these data until you are certain that the instrument performs satisfactorily.
EMC considerations
Canada spectrum management emissions notice
This class A digital product apparatus complies with Canadian ICES-001.
Cet appareil numérique de la classe A est conforme à la norme NMB-001.
ISM Classifickation: ISM Group 1 Class A
This classification has been assigned in accordance with CISPR 11 Industrial Scientific and
Medical, (ISM) instrument requirements.
Group 1 products apply to intentionally generated and/or used conductively coupled
radio-frequency energy that is necessary for the internal functioning of the equipment.
Class A products are suitable for use in all establishments other than residential locations and
those directly connected to a low voltage power supply network supplying a building for domestic
purposes.
There may be potential difficulties in ensuring electromagnetic compatibility in other environments
due to conducted as well as radiated disturbances.
Содержание ACQUITY QDa
Страница 2: ...June 28 2016 715003956 Rev C Page ii ...
Страница 16: ...June 28 2016 715003956 Rev C Page xvi ...
Страница 66: ...June 28 2016 715003956 Rev C Page 66 ...
Страница 70: ...June 28 2016 715003956 Rev C Page 70 ...
Страница 78: ...June 28 2016 715003956 Rev C Page 78 ...
Страница 104: ...June 28 2016 715003956 Rev C Page 104 ...