June 28, 2016, 715003956 Rev. C
Page 44
3.9.4
Fitting the ion guide assembly to the instrument
Required materials:
•
Chemical-resistant, powder-free gloves
•
2.5-mm Allen wrench
To fit the ion guide assembly to the instrument:
1.
Fit the differential aperture O-ring to the differential aperture support (see the figure on
).
2.
Fit the differential aperture to the differential aperture support using the 3 slotted screws,
and secure the screws using a flat-blade screwdriver.
3.
Carefully slide the ion guide’s PCBs into the differential aperture support.
4.
Secure the ion guide to the differential aperture support using the 4 ion guide securing
screws.
Figure 3–12:
Assembling the ion guide housing
5.
Secure the ion guide and differential aperture support assembly to the pumping block’s ion
guide support rods using the 2 differential aperture support screws, and tighten them using
the 3-mm Allen wrench.
Tip:
It is normal for the ion guide assembly to have some freedom to move when
assembled correctly.
6.
Carefully slide the pumping block and ion guide assembly into the instrument’s source
housing.
Tip:
To correctly orientate the pumping block, ensure that the ion guide remains visible at
the left-hand edge of the pumping block’s front face.
Warning:
To avoid personal contamination with biohazards or toxic materials, always wear
chemical-resistant, powder-free gloves while performing this procedure.
Содержание ACQUITY QDa
Страница 2: ...June 28 2016 715003956 Rev C Page ii ...
Страница 16: ...June 28 2016 715003956 Rev C Page xvi ...
Страница 66: ...June 28 2016 715003956 Rev C Page 66 ...
Страница 70: ...June 28 2016 715003956 Rev C Page 70 ...
Страница 78: ...June 28 2016 715003956 Rev C Page 78 ...
Страница 104: ...June 28 2016 715003956 Rev C Page 104 ...