June 28, 2016, 715003956 Rev. C
Page 97
Notes:
•
Performance QDa Detectors use a separate rotary backing pump, not an attached
diaphragm pump.
•
The 4-mm bulkhead fitting and O.D. drain-tubing is only used with “KAB” models, and
earlier.
3.
Secure the bottle-stopper onto the 2.5-L bottle (see the figure
, below).
Note:
The Performance QDa uses a rotary backing pump, which does not require the 6-mm
O.D. drain tube. To minimize unused tubing, cut the 3-m length of 6-mm O.D. tubing close
to the waste container.
4.
Position the bottle in a readily accessible location below the instrument.
Figure D–14:
2.5-L waste container tubing
D.8
Connecting the nitrogen exhaust line
Required materials
•
Chemical-resistant, powder-free gloves
•
Sharp knife
•
12-mm PTFE tubing (included in the Waters Rough Pump Connect Kit)
•
12-mm right-angle elbow connectors (included in the Waters Start-Up Kit)
Warning:
To prevent damage to the instrument, ensure that these criteria are met:
• The drain tubing is not submerged in solvent within the waste container.
• The waste container is not used to collect solvent waste from other sources.
2.5-L bottle
Bottle-stopper
4-mm O.D. drain tubing
6-mm O.D. tubing
(3-m hose)