June 28, 2016, 715003956 Rev. C
Page 50
Figure 3–16:
Backing-pump rear panel
6.
Tilt the pump slightly, and catch the oil in a suitable container.
7.
Dispose of the oil according to local environmental regulations.
8.
Insert the oil-drain plug into the pump’s rear panel.
9.
To flush the pump, pour 50-mL of fresh oil into the pump inlet, on the pump’s top side.
Tip:
If you encounter difficulty, remove the separator.
10. Operate the pump briefly.
11. Drain the flushing oil into a suitable container.
12. If necessary, repeat steps 9 through 11 until all contaminants are removed.
13. Remove the oil-inlet plug, and pour fresh oil into the oil-inlet port until the level reaches the
sight glass’s “max” mark.
14. Insert the oil-drain plug into the pump’s rear panel.
Notice:
Do not overfill the pump. That is, do not fill it beyond the “max” mark.
Oil level sight glass