June 28, 2016, 715003956 Rev. C
Page 37
3.7.5
Refitting the source components
Required materials:
•
Chemical-resistant, powder-free gloves
•
2.5-mm Allen wrench
To refit the source components:
1.
If you removed the O-rings from the source components, refit them.
2.
Insert the rubber entrance-aperture seal into the sample cone.
3.
Insert the entrance-aperture carrier into the and sample cone assembly.
Note:
Ensure that the entrance aperture seal is in place in the entrance-aperture carrier.
4.
Insert the entrance-aperture carrier and sample cone assembly into the entrance-aperture
on the right-hand-side of the ion block.
5.
Insert the calibration pin into the top of the ion block.
6.
Place the ion block assembly against the ion block support on the front of the instrument,
aligning the screw positions.
7.
Secure the ion block to the instrument using the 2, 3-mm screws, and tighten the screws
using the 2.5-mm Allen wrench.
8.
Refit the source enclosure (see
3.8
Cleaning the instrument case
Use a soft cloth, dampened with water, to clean the outside surfaces of the mass spectrometer.
3.9
Cleaning the ion guide assembly
Clean the ion guide assembly when these conditions apply:
•
You have dismissed LC and sample-related causes for decreased signal intensity.
•
Cleaning the source components fails to increase signal stability.
•
Replacing the entrance-aperture seal and disc fails to increase signal stability.
Requirement:
Wear clean, chemical-resistant, powder-free gloves when connecting the
source exhaust valve assembly.
Notice:
To avoid abrading the surfaces of the instrument, do not use abrasives or solvents
when cleaning them.
Содержание ACQUITY QDa
Страница 2: ...June 28 2016 715003956 Rev C Page ii ...
Страница 16: ...June 28 2016 715003956 Rev C Page xvi ...
Страница 66: ...June 28 2016 715003956 Rev C Page 66 ...
Страница 70: ...June 28 2016 715003956 Rev C Page 70 ...
Страница 78: ...June 28 2016 715003956 Rev C Page 78 ...
Страница 104: ...June 28 2016 715003956 Rev C Page 104 ...