June 28, 2016, 715003956 Rev. C
Page 30
3.6.2
Fitting the source enclosure to the instrument
Required material
Chemical-resistant, powder-free gloves
To fit the source enclosure:
1.
Using both hands, slide the source enclosure onto the instrument’s supporting rods.
2.
To secure the enclosure against the instrument, tighten the 2 thumbscrews on the front of
the source enclosure.
3.
Connect the electrical cable to the socket on the right-hand side of the instrument’s front
panel, and tighten the screws.
4.
Fit the probe to the source enclosure (see
3.7
Maintaining the source components
Clean the source components when these conditions apply:
•
The sample cone and cone gas nozzle are visibly damaged.
•
You have dismissed inlet and sample-related causes for decreased signal intensity.
3.7.1
Removing the sample cone assembly from the ion block
Required material
Chemical-resistant, powder-free gloves.
Warning:
To avoid personal contamination with biohazards or toxic materials, wear
chemical-resistant, powder-free gloves while performing this procedure.
Notice:
To avoid damage to the electrical connector’s screws, do not overtighten
them.
Notice:
To avoid damaging the fragile probe, use care when inserting it into the
source enclosure.
Warning:
To avoid personal contamination with biohazards or toxic materials, always wear
chemical-resistant, powder-free gloves while performing this procedure.
Содержание ACQUITY QDa
Страница 2: ...June 28 2016 715003956 Rev C Page ii ...
Страница 16: ...June 28 2016 715003956 Rev C Page xvi ...
Страница 66: ...June 28 2016 715003956 Rev C Page 66 ...
Страница 70: ...June 28 2016 715003956 Rev C Page 70 ...
Страница 78: ...June 28 2016 715003956 Rev C Page 78 ...
Страница 104: ...June 28 2016 715003956 Rev C Page 104 ...