June 28, 2016, 715003956 Rev. C
Page 32
Figure 3–3:
Removing the sample cone assembly
Figure 3–4:
Sample cone assembly components
3.7.2
Replacing the entrance-aperture seal and disc
Replace the entrance seal aperture when these conditions apply:
•
You have dismissed LC and sample-related causes for decreased signal intensity.
•
Cleaning the source components fails to increase signal stability.
Sample cone assembly
Ion block
Calibration pin
Entrance-aperture carrier
Disc aperture
assembly
Sample cone