![Waters ACQUITY QDa Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/waters/acquity-qda/acquity-qda_overview-and-maintenance-manual_3680840049.webp)
June 28, 2016, 715003956 Rev. C
Page 49
Required materials:
•
Chemical-resistant, powder-free gloves
•
Tray on which to place the pump
•
Container to catch used oil
•
Suitable pump oil
To add oil to the backing pump:
1.
Power-off the instrument using the power button in the top, left-hand corner of the front
panel.
2.
Wait for approximately 5 minutes to allow the instrument to vent.
3.
Switch off the pump, and disconnect the power cable.
4.
Place the pump on a tray suitable for catching dripping oil.
5.
Remove the oil-drain plug from the pump’s rear panel.
Warning:
To avoid personal contamination with biohazards or toxic materials in used pump oil,
always wear chemical-resistant, powder-free gloves while performing this procedure.
Warning:
To avoid burn injuries, allow the pump to cool before touching surfaces
displaying the burn warning symbol.