Figure D–10:
Rear configuration with LC stack
Tip:
This figure shows a QDa Detector fitted with the external, source exhaust valve assembly
(serial number suffix "KAB"). To see the QDa Detector fitted with the internal, source exhaust
valve (serial number suffix "KAD"), see the figures on
(rear configuration) and
(front configuration).
D.7.2
Alternate configurations
You can adopt an exhaust configuration presented in this section if it better suits your needs than
the configurations detailed on
Notice:
To prevent solvent leaks and resultant damage to instruments whose waste lines
are configured according to any of the arrangements presented in this document, ensure
that the entire PTFE tube serving as the waste conduit between the instrument and the
exhaust-trap bottle follows a continuous, downward gradient, without loops or areas of
compression.
Source exhaust valve
assembly
90-degree elbow
To laboratory
exhaust system
Exhaust-trap bottle
ACQUITY QDa
PTFE tubing
Convoluted
liquid waste
tubing
Liquid waste
container
Exhaust solenoid cable
PTFE tubing (ensure
downward gradient)
Содержание ACQUITY QDa
Страница 2: ...June 28 2016 715003956 Rev C Page ii ...
Страница 16: ...June 28 2016 715003956 Rev C Page xvi ...
Страница 66: ...June 28 2016 715003956 Rev C Page 66 ...
Страница 70: ...June 28 2016 715003956 Rev C Page 70 ...
Страница 78: ...June 28 2016 715003956 Rev C Page 78 ...
Страница 104: ...June 28 2016 715003956 Rev C Page 104 ...