December 5, 2017, 715004753 Rev. B
Page 59
Lamp light flashes red and
power light constant green
Failed startup diagnostic tests
Reseat and check alignment
of flow cell (see
).
Flush the flow cell (see
Dirty flow cell causing shutter
diagnostic test to fail
Flush the flow cell (see
Insufficient energy reaching
photodiode array because of
air bubble
Flush the flow cell (see
), or apply slight
backpressure of 207 to
345 kPa (2 to 3 bar, 30 to
50 psi) on the detector waste
outlet. For example, connect a
1- to 2-foot (30- to 60-cm)
length of 0.009-inch (0.23-mm)
ID tubing connected to the
detector waste outlet.
Weak lamp
Replace the lamp (see
Shutter failure message
Shutter failed
1. Clear bubbles from the flow
cell (see
2. Cycle power to the
detector.
Solvent in drain line
Leak from flow cell gasket
Replace the flow cell (see
Leak from flow cell inlet and
outlet fittings
Check fittings for
overtightening or
undertightening, and replace
fittings if necessary.
Table 4–3:
2998 PDA detector troubleshooting (continued)
Symptom
Possible cause
Corrective action