December 5, 2017, 715004753 Rev. B
Page 57
4.2.2.2
Lamp LED
The lamp LED, to the left of the power LED, indicates the lamp status.
4.2.3
Clearing bubbles from the flow cell
Poor energy causing lamp lighting failures or changes in the reference spectrum can cause
bubbles in the flow cell. The bubbles can cause baseline anomalies such as spiking and drift.
To clear bubbles from the flow cell:
Establish a flow of degassed acetonitrile or methanol through the flow cell at the rate
anticipated for the subsequent analysis.
Table 4–2:
Lamp LED indications
LED mode and color
Description
Unlit
Indicates that the detector lamp is extinguished.
Constant green
Indicates that the detector lamp is ignited and that the detector
passed the sequence of startup diagnostic tests.
Flashing green
Indicates that the detector is initializing or verifying wavelength
calibration.
Flashing red
Indicates that an error stopped the detector. The console's log
records information about errors that cause operational failure
See also:
2998 PDA Detector Online Help
Constant red
Indicates a detector failure that prevents further operation.
Cycle power to the detector. If the LED is still steady red,
contact your Waters service representative.
Содержание 2998
Страница 2: ...December 5 2017 715004753 Rev B Page ii...
Страница 8: ...December 5 2017 715004753 Rev B Page viii...
Страница 14: ...December 5 2017 715004753 Rev B Page xiv...
Страница 26: ...December 5 2017 715004753 Rev B Page 26...
Страница 44: ...December 5 2017 715004753 Rev B Page 44...
Страница 60: ...December 5 2017 715004753 Rev B Page 60...
Страница 68: ...December 5 2017 715004753 Rev B Page 68...