December 5, 2017, 715004753 Rev. B
Page 35
To connect the detector to other devices, use the two analog-out/event-in (I/O) connectors and
their mating connectors on the rear panel.
2.6.3.1
Generating an inject-start function
To generate the detector’s inject-start function at the start of an injection, connect cables as shown
in the following table and figure.
Tip:
The firmware defaults to Autozero on injection.
Figure 2–8:
Inject-start connections between the Alliance separations module and the
2998 PDA detector
2.6.3.2
Generating a stop-flow signal
The detector is fitted with a programmable switch output controlled by threshold or timed events.
To enable the generation of a stop flow, connect cables shown in the following table and figure.
Requirement:
To automatically stop the solvent flow within the system in the event of an error
condition or hardware failure, the stop-flow signal must be sent to the system’s pump.
Table 2–4:
2998 PDA detector connections to an Alliance separations module
Alliance separations module (connector B)
2998 PDA detector (connector A)
Pin 1 inject start (red)
Pin 1 inject start
(red)
Pin 2 inject start (black)
Pin 2 inject start
(black)
Red
Black
1
2
3
4
5
6
7
8
9
10
Inject Start
Inject Start
Ground
Lamp On/Off
Lamp On/Off
Chart Mark
Chart Mark
Ground
Auto Zero
Auto Zero
1
2
3
4
5
6
7
8
9
10
11
12
Inject Start
Inject Start
Ground
Stop Flow
Stop Flow
Hold Inject 1
Hold Inject 1
Hold Inject 2
Hold Inject 2
Ground
Chart Out
Chart Out
Alliance connector B
2998 PDA detector connector A
Содержание 2998
Страница 2: ...December 5 2017 715004753 Rev B Page ii...
Страница 8: ...December 5 2017 715004753 Rev B Page viii...
Страница 14: ...December 5 2017 715004753 Rev B Page xiv...
Страница 26: ...December 5 2017 715004753 Rev B Page 26...
Страница 44: ...December 5 2017 715004753 Rev B Page 44...
Страница 60: ...December 5 2017 715004753 Rev B Page 60...
Страница 68: ...December 5 2017 715004753 Rev B Page 68...