December 5, 2017, 715004753 Rev. B
Page 47
Tip:
Conditions other than a dirty flow cell can cause decreased lamp intensity. For more
information, refer to
Maintaining the flow cell consists of these tasks:
•
Flushing
•
Removing
•
Disassembling and cleaning
•
Installing a replacement
3.4.1
Flushing the flow cell
Required tools and materials
•
HPLC-grade water
•
HPLC-grade methanol
If the flow cell requires cleaning, first flush it with solvent.
To flush the flow cell:
1.
Select a solvent compatible with the samples and mobile phases that you have been using.
If you have been using buffers, flush the cell with 10 mL of HPLC-grade water and then
flush with 10 mL of a low-surface-tension solvent, such as methanol.
Requirement:
Ensure that the solvent is miscible with the previous mobile phase.
2.
Test the energy by performing the Read energy diagnostic test (see
).
Note:
If the diagnostic test fails and the has not been used more than 2000 hours or 1 year
from date of purchase (whichever event occurs first), call Waters Technical Service (see
).
3.4.2
Replacing the flow cell
Required tools and materials
•
1/4-inch flat-blade screwdriver
•
Flow cell
To replace the flow cell:
1.
Power-off the detector.
2.
Stop the solvent flow.
3.
Open the door.
4.
Disconnect the column-outlet tubing from the flow cell inlet.
5.
Disconnect the waste line from the flow cell outlet.
Содержание 2998
Страница 2: ...December 5 2017 715004753 Rev B Page ii...
Страница 8: ...December 5 2017 715004753 Rev B Page viii...
Страница 14: ...December 5 2017 715004753 Rev B Page xiv...
Страница 26: ...December 5 2017 715004753 Rev B Page 26...
Страница 44: ...December 5 2017 715004753 Rev B Page 44...
Страница 60: ...December 5 2017 715004753 Rev B Page 60...
Страница 68: ...December 5 2017 715004753 Rev B Page 68...