December 5, 2017, 715004753 Rev. B
Page 36
Figure 2–9:
Stop-flow connections between the Alliance separations module and the
2998 PDA detector
2.6.3.3
Connecting to an Empower or MassLynx data system via an eSAT/IN
module
To send an integrator analog output signal (–0.1 to
2.1 V) from the 2998 PDA detector to an
Empower or MassLynx system (via a two-channel eSAT/IN module), connect cables shown in the
following table and figure.
Table 2–5:
2998 PDA detector connections to an Alliance separations module
Alliance separations module (connector B)
2998 PDA detector (connector B)
Pin 4 stop flow
(red)
Pin 6 switch 1
(red)
Pin 5 stop flow – (black)
Pin 7 switch 1 – (black)
Table 2–6:
2998 PDA detector connections to the eSAT/IN module
eSAT/IN module connector
2998 PDA detector (connector B)
CHANNEL 1
Pin 1 signal out
(white)
Pin 2 signal out – (black)
Red
Black
1
2
3
4
5
6
7
8
9
10
Analog 1
Analog 1
Ground
Analog 2
Analog 2
Switch 1
Switch 1
Ground
Switch 2
Switch 2
Inject Start
Inject Start
Ground
Stop Flow
Stop Flow
Hold Inject 1
Hold Inject 1
Hold Inject 2
Hold Inject 2
Ground
Chart Out
Chart Out
1
2
3
4
5
6
7
8
9
10
11
12
Alliance connector B
2998 PDA detector connector B
Содержание 2998
Страница 2: ...December 5 2017 715004753 Rev B Page ii...
Страница 8: ...December 5 2017 715004753 Rev B Page viii...
Страница 14: ...December 5 2017 715004753 Rev B Page xiv...
Страница 26: ...December 5 2017 715004753 Rev B Page 26...
Страница 44: ...December 5 2017 715004753 Rev B Page 44...
Страница 60: ...December 5 2017 715004753 Rev B Page 60...
Страница 68: ...December 5 2017 715004753 Rev B Page 68...