1.
Power-off the detector, and disconnect the power cord from the power-entry module.
Figure 3–4:
Power-entry module
2.
Pinch the sides of the spring-loaded fuse holder, which is below the power entry module on
the rear panel of the detector.
Figure 3–5:
Sides of spring-loaded fuse holder
3.
With minimum pressure, withdraw the spring-loaded fuse holder.
Figure 3–6:
Removing fuse holder
4.
Remove and discard the fuses.
5.
Ensure that the new fuses are properly rated for your requirements, and then insert them
into the holder and the holder into the power entry module, gently pushing until the
assembly locks into position.
6.
Reconnect the power cord to the power entry module.
Power-entry module
Sides of fuse holder
Fuse holder
Power-entry module
Fuse (2)