2-34
Installing the Detector
Auto Zero connections
To auto zero the detector at the start of an injection, make the connections
described in the table and illustrated in the figure below. Use any available
pair of Inject Start terminals on the autosampler.
Autosampler Auto Zero connections to the detector
Autosampler and detector Auto Zero connections
2489 Detector
(A Inputs)
717plus
Autosampler
Terminal
Pin 9 Auto Zero +
Inject Start + (any
one of three paired
with –)
Pin 10 Auto Zero –
Inject Start – (any
one of three paired
with +)