3-64
Preparing the Detector
2.
Press Lamp (Shift 1) again to turn the lamp off. The absorbance screen
appears with an X through the lamp indicator icon, and the display
shows the words “Lamp off.”
Lamp off/on sequence
To turn on the lamp manually:
1.
When there is an X through the lamp icon on the absorbance screen,
press Lamp (Shift 1). The lamp control screen appears again with 0
hours and 00 minutes in the “Lamp has been on” field.
2.
Press Lamp (Shift 1) again to turn the lamp on. The absorbance screen
appears with the message “Igniting.” The lamp can take up to 1 minute
to illuminate. Once the lamp illuminates, the display returns to the
absorbance screen and the X is removed from the lamp icon.
You can conserve lamp life by programming it to turn on and off
automatically, for example, overnight, using a timed event method.
To program the lamp to turn on or off, you can select the Timed events option
in the Method choice list, or program the lamp through one of the external
contact closures.
•
See
“Programming timed events, threshold events, and methods”
, and the table on
for more information on
programming the lamp to turn on or off using a timed event.
•
See
“Configuring event inputs (contact closures)”
for more
information on programming the lamp through the external contact
closure.
Lamp off indicator
Lamp on indicator