2-32
Installing the Detector
Pump Chart Mark connections
Inject Start connections
Tip:
If the detector is connected to an Empower data system, use the Inject
Start connections to initiate the start of data acquisition.
To make Inject Start connections between the pump and the detector to start
a method:
1.
Using a signal cable, make the connections between the detector and the
pump that are summarized in the table and illustrated in the figure
below.
2.
Program the pump to provide a pulse output on the selected switch at
the beginning of each run. Refer to the Waters 600E Multisolvent
Delivery System User’s Guide for more details.
II
Waters 2489
TP01491
INJECT
STOP
FLOW
HOLD
S1
S2
S3
S4
SWITCHES
CHART
+
PRESSURE
+
CHART
_
PRESSURE
_
GND
GND
GND
GND
AUX.
+
12V