Connecting to other devices
2-35
Inject Start connections
You can also use the Inject Start connections on the autosampler to connect to
the Inject Start signals on the detector to program the start of the active
method.
To program an inject start, make the connections described in the table and
illustrated in the figure below. Use any available pair of Inject Start terminals
on the autosampler.
Detector Inject Start connections
Autosampler and detector Inject Start connections
2489 Detector
(II)
717plus Autosampler
Terminal
Pin 1 Inject Start + Inject Start + (any one of three
paired with
+
)
Pin 2 Inject Start –
Inject Start – (any one of three
paired with – )
II
Waters 2489
123456
7
8
9
10
11
12
13
14
15
TP01522A
Waters 717plus
Содержание 2489
Страница 8: ...viii...
Страница 32: ...1 16 Theory and Principles of Operation...
Страница 134: ...3 66 Preparing the Detector...
Страница 193: ...Warnings that apply to all Waters instruments A 9...
Страница 204: ...B 6 Detector Specifications...
Страница 220: ...D 14 Solvent Considerations...