3-42
Preparing the Detector
Viewing events within a method
To view timed and threshold events that make up a stored method:
1.
Retrieve the method (see
“Retrieving a method” on page 3-41
). Once you
have entered the method number to retrieve, the Method choice list
appears with the method number displayed within the method number
icon.
2.
Press 1 to view the timed events or 2 to view threshold events within the
displayed method.
If you make a change to a timed or threshold event within a method, the
asterisk appears (Method *) indicating that the method (*) is no longer
the same as the stored method you retrieved in step 1. You can then
store the method containing the altered event(s) in the same storage
slot.
Resetting a method
Resetting a stored method is a two-step process. First, you reset the current
conditions to the defaults; then you save the defaults in one of the storage
locations. The table on
lists the parameter default settings.
To clear one or more methods:
1.
Return to the Method choice list by pressing METHOD (Shift A/B).
2.
Press 5, Reset method *.
A message screen appears, asking whether it is okay to set current
conditions to factory defaults.
Reset method message
Содержание 2489
Страница 8: ...viii...
Страница 32: ...1 16 Theory and Principles of Operation...
Страница 134: ...3 66 Preparing the Detector...
Страница 193: ...Warnings that apply to all Waters instruments A 9...
Страница 204: ...B 6 Detector Specifications...
Страница 220: ...D 14 Solvent Considerations...