Installing the Detector
Starting a method
To start a method on the detector at the start of an injection from the
separations module, make the connections summarized in the table below and
illustrated in the figure below.
2695 separations module connections to the detector for starting a method
Turning the detector lamp on or off
Before you can turn the detector lamp on or off from the separations module,
you need to configure the lamp on/off signal at the front panel. You must
change the default lamp configuration parameter from Ignore to High or Low.
See the discussion
“Configuring event inputs (contact closures)”
for more information.
2695 Separations Module
(B Inputs and Outputs)
2489 Detector (II)
Pin 1 Inject Start
Pin 1 Inject Start
+
Pin 2 Inject Start
Pin 2 Inject Start –
Waters 2489
II
TP01525
Waters 2695
B