Maintaining the flow cell
4-15
Replacing the flow cell
The detector is shipped with a standard analytical flow cell installed. Replace
the flow cell when
•
the flow cell becomes damaged.
•
you want to use one of the optional flow cells (see the table on
To prepare for replacing the flow cell:
1.
Unpack and inspect the new flow cell.
2.
Power-off the detector and disconnect the power cord.
3.
Remove the front left panel cover.
4.
Disconnect the detector inlet/outlet tubing from the main column
connection and cap.
To replace the flow cell:
1.
Using a 1/4-inch flat-blade screwdriver, loosen the three captive screws
on the flow cell assembly front plate (see the figure on
).
2.
Gently pull the assembly toward you.
3.
Insert the new flow cell assembly into the detector.
4.
Tighten the captive screws.
5.
Reconnect the inlet/outlet tubing to your LC system.
6.
Reconnect the power cord, and then power-on the detector.
Caution:
To prevent contamination, use powder-free finger cots or
gloves when disassembling, inspecting, cleaning, or replacing parts
within the Waters TaperSlit Flow Cell or when removing or
replacing the flow cell within its assembly.