June 11, 2015, 715004754 Rev. A
Page 69
5.2 Operating in RIU mode
RIU mode provides front-panel method programmability. RIU mode sets µRIU as the
parameter displayed on the front panel and generated at the detector output, using
the RIU-FS parameter to convert RIU to a voltage.
Alternative:
RIU mode can also be controlled using chromatography data software.
5.2.1 Using the display in RIU mode
The detector provides a 128 × 64 bitmap graphic display and a 24-key membrane
keypad as the operator interface. After the startup diagnostics run successfully, the
detector displays the home screen.
Figure 5–5:
Home screen RIU mode
You can recall the home screen from any page by pressing the HOME key. At the first
use of the detector, the home screen shows the factory-set defaults. After the first
use, the home screen shows the settings displayed before the detector was last
powered-off.
5.2.1.1
Refractive index, function, and message icons
The HOME, TEMP °C, and CONFIGURE screens in the detector program display the
icons or fields shown in the figure above.
Table 5–2:
Mode, function, and message screen icons
Icon or field
Icon/field name
Function
Home1(status),
Home2(entry)
RIU-FS
Shows the current RIU-FS setting.
Home1(status),
Temp °C (entry)
Detector Temperature
Shows the current temperature of
detector oven.
Numerical field
RIU
Displays current refractive index
in µRIU (1 × 10
-6
refractive index
units).
Purge
RIU full scale
Detector temperature
Refractive index units
Polarity
Mode
Shift
Keypad lock/unlock
Local (method #)/remote control
Run time (minutes)
Next screen
Содержание 2414
Страница 2: ...June 11 2015 715004754 Rev A Page ii...
Страница 14: ...June 11 2015 715004754 Rev A Page xiv...
Страница 42: ...June 11 2015 715004754 Rev A Page 42...
Страница 56: ...June 11 2015 715004754 Rev A Page 56...
Страница 62: ...June 11 2015 715004754 Rev A Page 62...
Страница 92: ...June 11 2015 715004754 Rev A Page 92...
Страница 114: ...June 11 2015 715004754 Rev A Page 114...
Страница 125: ...June 11 2015 715004754 Rev A Page 125 Lower limit of temperature Temperature limitation Symbol Description...
Страница 126: ...June 11 2015 715004754 Rev A Page 126...
Страница 130: ...June 11 2015 715004754 Rev A Page 130...