June 11, 2015, 715004754 Rev. A
Page 37
2.4 Stacking system modules
This procedure applies to system modules that are equipped with interlocking
features.
To stack the modules:
1. Install two alignment pins to the bottom of the module that will be placed on
top.
2. Place the rear feet of the module that you are adding atop the previously added
module in the system stack, and slide it backward until its rear alignment pin
rests in the rear alignment slot on the previously added module.
Figure 2–1:
Aligning pins with slots
3. Lower the front of the module that you are adding so that its front alignment pin
rests in the front alignment slot on the previously added module.
4. Repeats steps 1 and 2 for the remaining system modules.
2.5 Connecting to the electricity source
The RI detector requires a separate, grounded power source. The ground connection
in the power outlet must be common and physically close to the module.
Warning:
To avoid spinal and muscular injury, do not attempt to lift a module
without assistance.
Warning:
To avoid crushing your fingers beneath or between modules, use
extreme care when installing a module in the system stack.
Alignment pin (2)
Alignment slot (2)
Содержание 2414
Страница 2: ...June 11 2015 715004754 Rev A Page ii...
Страница 14: ...June 11 2015 715004754 Rev A Page xiv...
Страница 42: ...June 11 2015 715004754 Rev A Page 42...
Страница 56: ...June 11 2015 715004754 Rev A Page 56...
Страница 62: ...June 11 2015 715004754 Rev A Page 62...
Страница 92: ...June 11 2015 715004754 Rev A Page 92...
Страница 114: ...June 11 2015 715004754 Rev A Page 114...
Страница 125: ...June 11 2015 715004754 Rev A Page 125 Lower limit of temperature Temperature limitation Symbol Description...
Страница 126: ...June 11 2015 715004754 Rev A Page 126...
Страница 130: ...June 11 2015 715004754 Rev A Page 130...