2.6 Connecting tubing and attachments
This section describes procedures for connecting the detector to these components:
•
A column or another detector
•
A waste container
•
The drip tray
The tubing connections for the detector are behind the front door and to the left of the
keypad.
Figure 2–3:
Tubing connections
2.6.1 Connect a column or second detector
Use only the insulating sample inlet tubing assembly supplied in the startup kit to
connect a column or second detector to the detector. This tubing assembly has been
Warning:
To avoid the harmful effects of personal contact with
solvents, including inhalation, always wear chemical-resistant,
powder-free gloves and observe good laboratory practice when you
handle them. See the Material Safety Data sheets for the solvents
you use.
Warning:
To avoid the harmful effects of personal contact with
leaking solvents, flush the detector with an appropriate miscible
solvent prior to installing or replacing fittings.
Warning:
To avoid hazards associated with spilled chemicals,
ensure waste containers are large enough to accommodate analysis
requirements.
Solenoid
purge valve
Inlet line
Recycle
valve
Pressure
relief valve
Содержание 2414
Страница 2: ...June 11 2015 715004754 Rev A Page ii...
Страница 14: ...June 11 2015 715004754 Rev A Page xiv...
Страница 42: ...June 11 2015 715004754 Rev A Page 42...
Страница 56: ...June 11 2015 715004754 Rev A Page 56...
Страница 62: ...June 11 2015 715004754 Rev A Page 62...
Страница 92: ...June 11 2015 715004754 Rev A Page 92...
Страница 114: ...June 11 2015 715004754 Rev A Page 114...
Страница 125: ...June 11 2015 715004754 Rev A Page 125 Lower limit of temperature Temperature limitation Symbol Description...
Страница 126: ...June 11 2015 715004754 Rev A Page 126...
Страница 130: ...June 11 2015 715004754 Rev A Page 130...