June 11, 2015, 715004754 Rev. A
Page 48
Requirement:
To meet the regulatory requirements of immunity from external
electrical disturbances that can affect the performance of this instrument, do not use
cables longer than 9.8 feet (3 meters) when you make connections to the
analog-out/event-in connectors. In addition, you must always connect the shield of
the cable to ground at one instrument only.
The following figure and table describe the two I/O connectors (and their
corresponding pin-outs) on the detector’s rear panel.
Figure 3–4:
Rear panel analog-out/event-in connectors
3.4.1 Connecting to a standalone Alliance e2695
separations module
When you use the Alliance e2695 separations module as a standalone controller (that
is, not on an Ethernet or IEEE-488 bus interface or under data system control), you
can make the following signal connections using the 2414 RI detector’s
analog-out/event-in connectors:
•
Autozero on inject
•
Chart mark on inject
•
Both chart mark and autozero on inject
•
Inject start
Table 3–3:
2414 RI detector analog-out/event-in connections
Signal Connections
Description
Chart Mark, Polarity and Polarity Enable,
Autozero, Purge, Recycle, and Inject Start
Accept TTL-level (0 to +5 V) or contact
closure signals from an external
instrument
Detector Out
Sends a ±2 V (full scale) signal to an
integrator or computer
Auxiliary Out
Sends a ±2 V (full scale) signal to an
integrator or computer (Temperature
Data)
B (inputs and outputs)
A (inputs and outputs)
1 + Autozero
2 - Autozero
3 GROUND
4 + Recycle Valve
5- Recycle Valve
6 + Polarity Enable
7 - Polarity Enable
8 GROUND
9 + Detector Out
10 - Detector Out
11 GROUND
12 + Auxiliary Out
13 - Auxiliary Out
1 + Inject Start
2 - Inject Start
3 GROUND
4 + Purge
5 - Purge
6 + Polarity
7 - Polarity
8 GROUND
9 + Chart Mark
10 - Chart Mark
11 GROUND
12 + Switch
13 - Switch
Содержание 2414
Страница 2: ...June 11 2015 715004754 Rev A Page ii...
Страница 14: ...June 11 2015 715004754 Rev A Page xiv...
Страница 42: ...June 11 2015 715004754 Rev A Page 42...
Страница 56: ...June 11 2015 715004754 Rev A Page 56...
Страница 62: ...June 11 2015 715004754 Rev A Page 62...
Страница 92: ...June 11 2015 715004754 Rev A Page 92...
Страница 114: ...June 11 2015 715004754 Rev A Page 114...
Страница 125: ...June 11 2015 715004754 Rev A Page 125 Lower limit of temperature Temperature limitation Symbol Description...
Страница 126: ...June 11 2015 715004754 Rev A Page 126...
Страница 130: ...June 11 2015 715004754 Rev A Page 130...