June 11, 2015, 715004754 Rev. A
Page 82
function screens of the home screen, press Next.
These are the secondary functions in
410 mode:
•
Sensitivity
•
Filter type
•
Time constant
•
Voltage offset
•
Enable/disable keypad entries
Figure 5–14: Secondary functions of the 410 mode home screen
From the home screen, you can access the following functions, either from the home
screen or by pressing the Next key:
•
Sensitivity– The Waters 410-compatible Integrator Output signal is defined by
the following relationship:
Detector Out = [
Δ
n/(5
×
10
-3
RIU/Volt)]
×
SENS
Sensitivity settings are 1, 2, 4, 8, 16, 32, 64, 128, 256, 512, and 1024 (default
is 4).