127
Type of Event
Status to generate a ‘1’ in the log Status to generate a ‘0’ in the log
Relay output
Relay activated
Relay not activated
Interlock DI
In the Interlock state as defined in
the Interlock page
In the non-Interlock state as defined in
the Interlock page
Level Switch DI
In the Low Alarm state as defined
in the Level Switches page
In the Normal state as defined in the
Level Switches page
Generic DI –
Alarm selected
In the Alarm state as defined in the
Generic Inputs page
In the Normal state as defined in the
Generic Inputs page
Generic DI – No
Alarm selected
Switch closed
Switch open
A window will open to ask if you would like to open the file or save it to disk. Select the method
you would like to use. The file is .csv (comma separated variable) file that will open in a
spreadsheet such as Excel.
6.37 Analog Output (1-4) Menu – Retransmit Mode
(only appears if one or more 4-20 mA output option cards are installed)
4-20 mA Output Control Mode
Click on the pull-down to select between Not Used , Proportional Feed and Retransmit control
mode. This section assumes that the control mode is set to retransmit.
Input Assignment
Click on the arrow to pull down the choices of which sensor measurement will be retransmitted as
a 4-20 mA signal. The choices are any of the sensor inputs, any of the sensor’s temperature
compensation inputs, LSI, RSI, any paddlewheel flow meter or any 4-20 mA input signal (only if
the 4-20 mA input option is installed).
Input Status
This column shows the status of the sensor that has been mapped to the 4-20 mA output. Possible
messages include Normal, Unmapped, and Sensor Error.
Input Reading
This column shows the process value of the input that is mapped to the 4-20 mA output. This will
be in units of conductivity, pH, ORP, temperature or whatever the 4-20 mA input is using for units
of measure.
Output mA Value
This column shows the status of the 4-20 mA output in mA.
Output Status
This column shows the status of the 4-20 mA output in %, as well as any error messages. It will
read Normal if the input value is between the two programmed values defining the 4-20 mA
output, or Over Range if the input is above the 20 mA value, or Under Range if it is below the 4
mA value.
Custom Name
Type the name you want to use to identify this output into the text box.
Содержание WebMaster