21
VoltageCode
R1 R2 R3 R4 R5 R6 R7 R8
0
or
4 DRY DRY DRY DRY DRY DRY DRY DRY
1 or 6
POWERED
POWERED
POWERED
POWERED POWERED POWERED POWERED DRY
2
or
5
POWERED POWERED POWERED POWERED POWERED POWERED POWERED POWERED
3 or 7
DRY
POWERED
DRY
POWERED
DRY POWERED
DRY POWERED
DRY CONTACTS WILL BE LABELED COM INSTEAD OF NC.
BLK
WHT
GRN
TO GND
STUD
WHT
BLK
GRN
TO GND
STUD
BLK
GRN
WHT
HO
T
GN
D
NE
U
GROUND
STUD
SOLENOID OR
MOTORIZED
BALL VALVE
METERING
PUMP
AC POWER
INPUT
NEU HOT
NC NO
NC NO
NC NO
NC NO
NC NO
NC NO
NC NO
NEUTRAL
SEE
DETAIL “A”
DETAIL “A”
BRACKET
PC
BOARD
SEPARATE
NUT
PRIMARY EARTH
CONNECTION
OTHER EARTH
CONNECTIONS
RELAY8 RELAY7 RELAY6 RELAY5 RELAY4 RELAY3 RELAY2 RELAY1
BLK
IF USING
A MOTORIZED
BALL VALVE
Important: The mains power supply earth ground connection
must be retained by separate nut.
Figure 4: Wiring Diagram for AC Power Input and Relay Outputs