28
4.3
Keypad
The keypad consists of 4 directional arrow keys (up, down, left and right), 4 menu/function keys
(previous, next, exit and enter), and 4 quick access keys (home, calibrate, relays, and help).
The arrow keys are used to change settings (up and down) and move the cursor to a different
location within that menu in order to change that setting (left and right).
The function keys are used to navigate through the different menu selections. The Next and Prev
keys move you from menu to menu. The Enter key is used to enter a submenu, and to accept a new
setting. The Exit key is used to back up one menu level. No matter where you are in the menus,
pressing Exit enough times will bring you to the Summary Screen (or press Home, see below).
To change a numeric value, move the cursor to the digit you want to change by pressing the left or
right arrow keys, then use the up or down arrow keys to change the value. Press the Enter key after
all digits have been changed to enter the new value into memory. Similarly, to change an
alphanumeric value, use the up or down arrow keys to change that character, and the left and right
arrow keys to move to the next character. Press the Enter key after all characters have been
changed.
The quick access keys bring you directly to frequently used or critical functions. The Home key
will bring you directly back to the Summary Screen. The CAL key will bring you directly to the
sensor calibration menus. The Relays key will bring you directly to menus that allow you to force
the relays off or on. The Help key is under construction. Refer to Section 5.0 for details on use of
the keypad for sensor calibration, relay control, setting communications parameters, etc.
4.4
Access Codes
The WebMasterONE series controller is shipped with the access codes at default values. In order
to prevent against unauthorized tampering, you will need to change the access codes to your own
values. See section 6.50 for details.
There are three levels of protection. If the
Master
password is used in the WebMasterONE
Authentication page, then any set point can be changed. The default Master password is 2001.
If the
Calibration
password is used in the Authentication page, then sensor calibrations will be
allowed, but other changes will not. The default Calibration password is 2002.
If the
Read Only
password is used the user will be able to look at any page, but will not be allowed
to make any changes. The default Read Only is 2003.
The default user name is "webmaster." This can and should be changed in the Access Code page.
See section 6.50 for details.
4.5 Initial
Startup
Turn the power switch on. Be patient, the start up sequence will take about the same time as it
takes a computer to boot up, approximately 2 minutes. At first the display backlight will blink and
all output LEDS will light. Around 30 seconds later, the display backlight will come on, and the
LEDS will shut off. Then the display will come up, and show you the % complete status of the
start up process. After around 1 minute, the Summary Screen will appear. In 15 seconds, you will
hear 3 beeps, and the controller is ready to go.
The WebMasterONE Series controllers have a wide range of capabilities, so the steps required
preparing the site for installation would be different depending upon the capability you will be
using.