Page 120 of 148
300004969171/B
Power Specifications—Input
Maximum Continuous
Power
550 watts
Hold-up Time
>40ms @ full load
Over voltage Protection
+15% of set voltage
Overload Protection
Auto Recover
Minimum Loading
None
Cross regulation
<0.5%
Line & Load regulation
<0.5%
Turn on Time
< 2 sec
EMC Compliance
Compliance to the following standards:
EN61000-4-2 ESD level3
EN61000-4-3 RF susceptibility 3v/m
EN61000-4-4 EFT level3
EN61000-4-5 Surge level3
EN61000-3-2 Harmonics
EN61000-3-3 Flicker
EN61000-4-11 Voltage dips, short interruptions
EN55011 conducted an radiated emissions
Fusing
The Isolation Transformer is merged with a fuse on each mains leg. The fuse values are
dependent on the transformer setting for the available AC mains power from the wall.
The transformer uses 5x20mm time delay fuses.
The AC mains input for the workstation power supply and 48VDC power supply are
internally fused and are not serviceable parts.
The 48VDC power supply output is protected with a 5A, 5x20mm time delay fuse.
Содержание CORE Mobile
Страница 19: ...300004969171 B Page 19 of 148 This page is intentionally blank for pagination purposes...
Страница 28: ...Page 28 of 148 300004969171 B Figure 10 Connector Panel...
Страница 38: ...Page 38 of 148 300004969171 B Figure 16 Settings VH IVUS Defaults...
Страница 43: ...300004969171 B Page 43 of 148 This page is intentionally blank for pagination purposes...
Страница 66: ...Page 66 of 148 300004969171 B This page is intentionally blank for pagination purposes...
Страница 85: ...300004969171 B Page 85 of 148 This page is intentionally blank for pagination purposes...
Страница 91: ...300004969171 B Page 91 of 148 This page is intentionally blank for pagination purposes...
Страница 93: ...300004969171 B Page 93 of 148 This page is intentionally blank for pagination purposes...
Страница 110: ...Page 110 of 148 300004969171 B This page is intentionally blank for pagination purposes...
Страница 116: ...Page 116 of 148 300004969171 B This page is intentionally blank for pagination purposes...