to assess quantitative blood flow or to develop a numeric measure of total cross-
sectional blood flow.
The ChromaFlo option can detect blood velocities in the following range:
o
The lower limit on particle detection is between 4 cm/sec and 7 cm/sec,
depending on the intervening attenuation. (The higher limit is obtained with
maximum tissue attenuation between the transducer and region of flow.)
o
The upper limit on particle detection is between 107 cm/sec and 110 cm/sec,
depending on the intervening attenuation. (The lower limit is obtained with
maximum tissue attenuation between the transducer and region of flow.)
Artifact
Cause
Resolution
Flow is not
ChromaFlo
detected
Velocity sensitivity of the ChromaFlo technique is
between 4 cm/sec and 7 cm/ sec, depending on the
attenuating tissue between the transducer and the
region of flow. The higher velocity number (lower
sensitivity) is obtained when detecting flow at
depths greater than 8mm. In either situation, flow
below this velocity will appear as black.
Carefully identify
and compare the
lumenal boundary
based on grayscale
images with those
obtained from
ChromaFlo images,
as well as other
modalities such as
angiography.
Reduced
frame rate
Due to processing speeds in ChromaFlo mode, the
overall frame rate of the Volcano system is lowered
to approximately 12 frames per second, which
represents combined grayscale and color features.
None. Very fast
moving structures
may appear to move
in a more continuous
manner during
standard, non-
ChromaFlo imaging.
Moving tissue
coded as color
The ChromaFlo technique uses the changing IQ
“speckle” pattern to detect regions of flow. It is
possible for tissue motions to generate speckle that
is detected as a flow region. The ChromaFlo
processor employs a sensitivity control which
discriminates between tissue and flow regions
based on the magnitude of the grayscale image.
For example, a bright reflector, even though it
generates an echo which is decoded as flow, is
suppressed. A lower level signal is not suppressed.
This is based on the clinical findings in which blood
Adjust the region of
interest as necessary
to prevent tissue
from being coded as
color.
Содержание CORE Mobile
Страница 19: ...300004969171 B Page 19 of 148 This page is intentionally blank for pagination purposes...
Страница 28: ...Page 28 of 148 300004969171 B Figure 10 Connector Panel...
Страница 38: ...Page 38 of 148 300004969171 B Figure 16 Settings VH IVUS Defaults...
Страница 43: ...300004969171 B Page 43 of 148 This page is intentionally blank for pagination purposes...
Страница 66: ...Page 66 of 148 300004969171 B This page is intentionally blank for pagination purposes...
Страница 85: ...300004969171 B Page 85 of 148 This page is intentionally blank for pagination purposes...
Страница 91: ...300004969171 B Page 91 of 148 This page is intentionally blank for pagination purposes...
Страница 93: ...300004969171 B Page 93 of 148 This page is intentionally blank for pagination purposes...
Страница 110: ...Page 110 of 148 300004969171 B This page is intentionally blank for pagination purposes...
Страница 116: ...Page 116 of 148 300004969171 B This page is intentionally blank for pagination purposes...