Setting Diameter
Diameter is the depth of the field of view to which the ultrasound image data is acquired.
The following table defines range of diameter and increments for each catheter.
Catheter Type
Diameter Range
Diameter Step Changes
Eagle Eye
Gold
8-20 mm
2
Eagle Eye
Platinum
8-20 mm
2
Pioneer Plus
8-20 mm
2
Revolution
8-14 mm
2
Visions PV .014P
8-20 mm
2
Visions PV .018
10-24 mm
2
Visions PV .035
20-60 mm
5
To change the diameter:
1
Open the Adjust Image dialog box.
Figure 23:
Adjust Image Dialog Box - Diameter
2
In the
Diameter
box, use the up and down arrows to increase or decrease the
Diameter setting. The IVUS image diameter will increase or decrease according to
the setting.
NOTE:
When the catheter is disconnected, the NearVu settings return to the default
values. Gain and Diameter settings persist and do not return to their default values.
EXCEPTION:
The Eagle Eye and VH Diameters do not return to default 10mm
diameter.
Содержание CORE Mobile
Страница 19: ...300004969171 B Page 19 of 148 This page is intentionally blank for pagination purposes...
Страница 28: ...Page 28 of 148 300004969171 B Figure 10 Connector Panel...
Страница 38: ...Page 38 of 148 300004969171 B Figure 16 Settings VH IVUS Defaults...
Страница 43: ...300004969171 B Page 43 of 148 This page is intentionally blank for pagination purposes...
Страница 66: ...Page 66 of 148 300004969171 B This page is intentionally blank for pagination purposes...
Страница 85: ...300004969171 B Page 85 of 148 This page is intentionally blank for pagination purposes...
Страница 91: ...300004969171 B Page 91 of 148 This page is intentionally blank for pagination purposes...
Страница 93: ...300004969171 B Page 93 of 148 This page is intentionally blank for pagination purposes...
Страница 110: ...Page 110 of 148 300004969171 B This page is intentionally blank for pagination purposes...
Страница 116: ...Page 116 of 148 300004969171 B This page is intentionally blank for pagination purposes...