Anti Crushing Verification and Adjustments
1.
Remove the cover of the Anticrushing PCB located at the back of the column.
2.
Measure with a multimeter between TP12 and TP9 (GND) and adjust to 4.5V with a flat
screwdriver in R3.
3.
Then measure with a multimeter between TP13 and TP9 (GND) and adjust to 0.5V with a flat
screwdriver in R4.
Generator J4 wiring
If J4 connector is not present generator interface will need to be hardwired.
Pin 1 - 3TS1-4 (Bucky Start)
Pin 2 - 3TS1-5 (Bucky Ready 1)
Pin 3 - 3TS1-7 (Ground)
Pin 4 - 3TS1-17 (Bucky Start 24vdc)
Troubleshooting
•
Verify U24 Chip on ATP Board is V7R1b39, If not call Viztek to obtain proper chips.
•
Verify Panel NIC has an IP of 192.168.95.1