Ultra Installation Guide v2.0
49
Customer Support
1.800.366.5343 –
VZ0022UG201509-2.0
17.
Using drop down menu Select the panel, by serial number, you want to calibrate.
18.
Select Connect
19.
Select To Calibration followed by the Calibration button
underneath it
20.
Select number of bright frames per column ( 4 is our
default)
21.
Select Offset
•
System will automatically collect offset images
22.
Take bright frames with about 20-30 seconds between
shots.
•
Set kvp, mAs, ma, and ms in tech service for shots.
•
All shots within each range must be made at same
kvp and mas
•
After each good shot, click next.