Ultra Installation Guide v2.0
48
Customer Support
1.800.366.5343 –
VZ0022UG201509-2.0
Panel calibration
This applies to ultra 3.3.0.6 or newer with Toshiba panel and driver 4001.
1.
Launch Ultra
2.
Click on Options
3.
Select device configuration tab
4.
Select Toshiba
5.
Scroll down to Exposure control
6.
Ensure share calibration is checked
7.
Exit ultra
8.
Launch Tech Service
9.
Launch Configuration on the right side
•
Find the work station using Bucky 1.
10.
Close Configuration.
11.
Launch Manual Calibration below configuration.
•
Enter the workstation number using Bucky 1 (usually WS1).
•
Enter User Mode
12.
Set the tube at a 72 inch SID
13.
Remove Grid
14.
Collimate to just outside the square on bucky’s face
15.
Insert 1.5 mm copper filter
16.
Open C:\opal\plugins64\Toshiba\TETD_FPD_Manager64.exe