Ultra Installation Guide v2.0
32
Customer Support
1.800.366.5343 –
VZ0022UG201509-2.0
1.
For Three Phase Generators, connect the Power wires L1, L2 and L3 to the Fuse Holders of F3, F4
and F5 (right side of the Cabinet), and the Ground wire to the Ground Studs in the Cabinet Frame
(located above these fuses or close to the right side of the HV Transformer).
2.
According to the nominal voltage of the line, verify or connect the wire “:” to the indicated
terminal (TB) of Transformer 6T2. This wire is factory connected to 230 VAC (for 1-Phase), 400
VAC (for 3-Phase) or 480 VAC / 530 VAC (for 80 kW 3-Phase Generators).
3.
Check 6T2-4 for low speed rotor only generators with Toshiba tubes is wired for 110vac/120vac.
•
When routing cables, take notice of cover mounting.