Ultra Installation Guide v2.0
39
Customer Support
1.800.366.5343 –
VZ0022UG201509-2.0
Figure 3
Figure 4
1.
Press Prep and Expose from the hand switch
. (Prep may take up to 45 seconds to start Calibration)
2.
“AutocalOK” will replace the
“AutoCalibration
” once the calibration is completed for the selected
focal spot.
3.
Spot check generator for calibration verification
. (E06 can be adjusted for proper cal)